Testing Procedures
Test procedures for PCB in Oil follow EPA SW846 Method 8082 and is described in current version of SOP GC-1.0 Analysis of Polychlorinated Biphenyls (PCBs) in Insulating Liquids by Gas Chromatography (GC)
Test procedures for Moisture Content in Oil are according to ASTM D 1533-00 and current version of SOP KF-1.0 Measurement of Water in Insulating Liquids by Coulometric Karl Fischer (KF).
Test procedures for Dielectric Breakdown Voltage are according to ASTM D 887-2, Procedure A and current version of SOP PR-5.0 Dielectric Breakdown Voltage of Insulating Liquids Using Disk Electrodes.

"Committed to maximizing returns on assets through versatility and innovation"